HCY Automatic Nucleic Acid Extraction System can be used in combination with viral DNA/RNA extraction reagents to enable nucleic acid extraction in a rapid, efficient and stable manner for subsequent clinical molecular diagnosis. The entire workflow can be executed in a biosafety cabinet to protect operators from potential hazards.
The System enables the parallel processing of up to 1-96 samples per run. With the pre-programmed protocols and magnetic bead-based reagent cartridges, the system provides consistent and stable nucleic acid purification for every busy laboratory.
Certifications: FDA/NMPA/RoHS
Product Name | Automatic Nucleic Acid Extractor |
Item No. | CY-NAAE-2106 |
Dimension (mm) | 600(L) × 420(W) × 500(H) |
Certifications | FDA / NMPA / RoHS |
Processing Volume | 20-400ul |
Extraction Throughput | 1-96, flexible |
Sample Types | cultured cells/tissues/buccal swabs/blood |
Well Plate | 96 deep well plate (V bottom) |
Consumables | 8-strip tip comb |
Nucleic acid extraction time | 0~30 minutes |
Heating temperature adjustable | From room temperature to 115°C/240°F |
Temperature control precision | ±0.1℃ |
Heating range | Room temperature +5℃~125℃ |
Reagent type | Magnetic bead-based, open-source reagents |
Nucleic acid extraction purity | A260/A280=1.6~1.9 |
Inter-well purification variation | CV<3% |
Magnetic beads recovery rate | ≥90% |
Touch screen size | A 7-inch touchable color screen |
Disinfection/decontamination method | UV lamp, 96-well prepacked plate and disposable tip |
Input power | AC 100-240V, 3.5-1.8A, 50/60Hz |
Product weight | 29±1k |
Benefits
- Decrease reagent cost over 83%
When sample quantity <16 pcs per run, only use 1 pc prefilled 96 Kit, can execute the nucleic acid extraction.
2.Flexible configuration of prefilled 96 Kit
Can adjust the reagent type & position in the 96 well plate, to match different extractor - Dismountable magnetic rod
When sample quantity per run<16pcs, or even <8 pcs, can put only 1 pc prefilled 96 Kit, and hand teardown the some magnetic rods. Then the same prefilled 96 Kit can be used in next extraction process. save much reagent cost.
4.High recovery of magnetic beads
After magnetic beads are absorbed on the head of the magnetic rod, the magnetic rod can be be lifted higher
0.1-0.2mm, for higher recovery of magnetic beads ensures high yield of nucleic acid. - New strong magnetic rod
With the newly designed strong magnetic rod, the magnetic beads are absorbed on the head of the magnetic rod, so as to ensure that the elution buffer can still cover all the magnetic beads even with a small elution volume.
6.Accurate heating
4 heating modules in each Unit, precise heating the stipulated well,avoid heating the adjacent - Precise movement of magnetic Rod
The magnetic rod is driven by Double-nut ball screw, to ensure that the rod runs smoothly, high precision and longer service life of the rod. - large magnetic rod vibration amplitude.
The driving device is equipped with a high-performance stepping motor, with large magnetic rod vibration amplitude. The vibration amplitude can be set according to the volume of the solution to ensure good and even blending effect. Each moving component is protected by the limit position protection mechanism to avoid instrument failure.
Consumables for the extraction system
- 96-well prepacked plate and
- 8-well regent tube
- Magnetic rod sleeve
Related Products